FDA Recall Terminated

Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.

Recall: Z-2057-2015 · Initiated March 24, 2015

Recall

Recall Number
Z-2057-2015
Event Number
71234
Firm
Hamilton Medical, Inc.
FEI Number
2937708
Product Code
CBK
Status
Terminated
Root Cause
Device Design
Initiated
March 24, 2015
Posted
July 10, 2015
Terminated
December 23, 2015
Address
4990 Energy Way, Reno, NV, 89502-4123

Description

Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.

Reason

Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.

Action

*** Please note *** This is not a new recall. An error occurred, and this recall was originally classified as a class I on July 10, 2015. No further action is required. Hamilton Medical AG notified their customers with a letter dated February 19, 2015 and had a fix (SW 2.42) that was available as of February 20th 2015. Hamilton Medical proceeded to send out notification letters the 24th of March 2015 once they compiled the complete and validated list of customers that had the affected software. Customers with questions were instructed to call 800-426-6331, ext.215.

Distribution

Nationwide Distribution

Quantity

248