FDA Recall Terminated

Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.

Recall: Z-2053-2020 · Initiated April 13, 2020

Recall

Recall Number
Z-2053-2020
Event Number
85147
Firm
Philips North America, LLC
FEI Number
1218950
Product Code
MKJ
Status
Terminated
Root Cause
Process control
Initiated
April 13, 2020
Terminated
April 7, 2021
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.

Reason

The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.

Action

On April 13, 2020, the firm distributed URGENT Medical Device Recall letters to affected customers. Customers were informed of the device defect, which may result in the following unexpected behavior: - The device may not turn on. - The device may not perform the selected function. - The device may deliver a shock with an energy level different from the setting selected by the user. There have been three reported patient deaths potentially associated with the failure of the HeartStart XL Monitor/Defibrillator rotary therapy selector switch. Customers were advised that the HeartStart XL Defibrillator/Monitor has been discontinued and has reached its end of life. Therefore, custoemrs should replace and retire their units as soon as practically possible. Customers should continue to perform Shift Checks and Operational checks as recommended in the Instructions for Use (IFU) as this reduces the risk of a failure until your units are safely retired. If you identify a device that exhibits any of the behaviors described above, or which fails shift or operational checks, immediately remove it from service. Customers were asked to complete and fax or email the included customer reply form to Philips. If you need further information or support concerning this notification, please contact your local Philips representative or call us at 1-800-722-9377.

Distribution

Worldwide distribution - US Nationwide distribution.

Quantity

94,034 units