FDA Recall Terminated

SHAPE ARM, DUAL SH2, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400254 and the Arm part number in the kit is P17084 . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.

Recall: Z-2053-2014 · Initiated March 17, 2014

Recall

Recall Number
Z-2053-2014
Event Number
67813
Firm
Stryker Communications
FEI Number
3002975342
Product Code
BRY
Status
Terminated
Root Cause
Process control
Initiated
March 17, 2014
Posted
July 15, 2014
Terminated
May 15, 2015
Address
1410 Lakeside Pkwy, Ste 100, Flower Mound, TX, 75028-4026

Description

SHAPE ARM, DUAL SH2, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400254 and the Arm part number in the kit is P17084 . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.

Reason

During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.

Action

Direct accounts in the US were notified by letter dated March 21, 2014, and asked to report any malfunctions to Stryker Communications immediately. In addition, Stryker Representatives will visit all US accounts to inspect all the product units on hand. If needed, Representatives will check that black, screw/lock washers supplied by Stryker are secured in place and will replace any that are not the typed furnished by Stryker. All units will be updated with a yellow, warning sticker on the monitor bracket which reads: "Failure to properly install the screw and lock washer below can result in the attached component/equipment falling." Stryker Communications' foreign distributors will oversee notification, inspection, and corrections of their accounts and report corrections and other findings to Stryker Communications.

Distribution

Worldwide Distribution. US nationwide including Puerto Rico, China, Philippines, Mexico, Korea, Japan, Colombia, Chile, Canada, Brazil, and Australia

Quantity

953 total STRYKER SHAPE Arms of all types