FDA Recall Terminated

Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit NL950-V and Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit with Ventricular Drainage System NL950-VC. This intracranial pressure (ICP) catheter is indicated for use on patients that require continuous invasive intracranial pressure monitoring and/or cerebrospinal drainage. Access is achieved through a tunneled technique using the included trocar.

Recall: Z-2053-2009 · Initiated March 16, 2009

Recall

Recall Number
Z-2053-2009
Event Number
51674
Firm
Integra Neuro Sciences
FEI Number
2023988
Product Code
JXG
Status
Terminated
Root Cause
Process design
Initiated
March 16, 2009
Posted
September 8, 2009
Terminated
July 19, 2012
Address
5955 Pacific Center Blvd, San Diego, CA, 92121

Description

Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit NL950-V and Integra NeuroSciences Ventrix Ventricular Tunneling Pressure Monitoring Kit with Ventricular Drainage System NL950-VC. This intracranial pressure (ICP) catheter is indicated for use on patients that require continuous invasive intracranial pressure monitoring and/or cerebrospinal drainage. Access is achieved through a tunneled technique using the included trocar.

Reason

Due to a dimensional error, the diameter of the stylet provided with the device may be too large to insert into the lumen of the catheter tip.

Action

An "Urgent-Product Recall Notification" letter dated March 23, 2009 was issued to consignees. The letter described the issue and instructed customers to examine their inventory immediately to determine whether they have any NL950V - Ventrix Ventricular Tunneling Pressure Monitoring Kits, or NL950VC - Ventrix Ventricular Tunneling Pressure Monitoring Kits with Ventricular Drainage System. Customers were further instructed not to use the above recalled product. Call Integra Customer Service at 1-800-654-2873, and then select 5 for 'Returns and Repairs' to obtain a Returned Material Authorization number, and return the product as directed. Direct questions about the recall to Integra NeuroSciences by calling 1-858-455-1115, extension185.

Distribution

Worldwide Distribution -- US (states of AR, CA, OH, TX, CT, NY, LA, FL, NY, IL, and VA), Belgium and Malaysia.

Quantity

274 kits