Boston Scientific's LATITUDE Patient Management System: LATITUDE 6.1 Patient Management System Website Server software, Model 6488 for United States Server software, Model 6465 for European Union LATITUDE 6.2 Patient Management System Website Server software, Model 6441 for Canada Server software, Model 6442 for Australia / New Zealand Intended for use to remotely communicate with a compatible Boston Scientific pulse generator and transfer data to a central database.
Recall
- Recall Number
- Z-2052-2011
- Event Number
- 57744
- Firm
- Boston Scientific CRM Corp
- FEI Number
- 2124215
- Product Code
- LWS
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 10, 2011
- Posted
- April 26, 2011
- Terminated
- November 15, 2012
- Address
- 4100 Hamline Ave N, Saint Paul, MN, 55112-5700
Description
Boston Scientific's LATITUDE Patient Management System: LATITUDE 6.1 Patient Management System Website Server software, Model 6488 for United States Server software, Model 6465 for European Union LATITUDE 6.2 Patient Management System Website Server software, Model 6441 for Canada Server software, Model 6442 for Australia / New Zealand Intended for use to remotely communicate with a compatible Boston Scientific pulse generator and transfer data to a central database.
Boston Scientific has determined that data within certain printed reports obtained from the LATITUDE Patient Management System may not be accurate, in that, the Most Recent Presenting Electrogram (MRP-EGM) for some of the selected patients may be either missing or incorrect.
Consignees were sent a Boston Scientific "LATITUDE Medical Records Information" letter dated January 2011. The letter was addressed to "Dear Doctor". The letter described the problem and the product involved in the recall. The letter also included the Rate of Occurrence, Action and Further Information sections.
Worldwide Distribution -- USA, including DC and PR, and countries of AUSTRALIA, CANADA, AUSTRIA, BELGIUM, DENMARK, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, NETHERLANDS, NORWAY, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, and UNITED KINGDOM.
25,836