NAVIS-EX , Image Filing Software for NIDEK AFC Fundus camera, Ver 1.1.0 to 1.3.5. NAVIS-EX is indicated for use as file management software for ophthalmic images.
Recall: Z-2046-2013
·
Initiated August 1, 2013
Recall
- Recall Number
- Z-2046-2013
- Event Number
- 65865
- Firm
- Nidek Inc
- FEI Number
- 2936921
- Product Code
- HKI
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- August 1, 2013
- Posted
- August 23, 2013
- Terminated
- February 25, 2014
- Address
- 47651 Westinghouse Dr, Fremont, CA, 94539-7474
Description
NAVIS-EX , Image Filing Software for NIDEK AFC Fundus camera, Ver 1.1.0 to 1.3.5. NAVIS-EX is indicated for use as file management software for ophthalmic images.
Reason
Under certain circumstances, the Left eye image may be saved as the Right eye image.
Action
"Medical Device Correction Notification" letters dated August 5, 2013 were sent to consignees through their US contracted distributor - Marco Opthalmic. The letter described the product issue and provided information on the correction. Customers can contact 800-223-9044 for questions.
Distribution
Nationwide US distribution only.
Quantity
86