nordicBrainEX Product Usage: nordicBrainEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities.
Recall
- Recall Number
- Z-2045-2018
- Event Number
- 80162
- Firm
- NordicNeuroLab AS Mollendalsveien 65c Bergen Norway
- FEI Number
- 3006738448
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- December 21, 2012
- Terminated
- September 16, 2020
Description
nordicBrainEX Product Usage: nordicBrainEx is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities.
There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following situations: A group of fibers has been selected using the VOI functionality and put into a fiber group. This is done with images series A visualized in the 3D viewer. Then a new images series B is visualized in the 3D viewer. If image series A and B have the same geometrical resolution (pixel sizes and slice distances), the coregistration matrix taking A to B will not be applied to the fiber group, and therefore the fiber group will in general not be positioned correctly on B in the 3D viewer. Furthermore, if the fiber group is exported as a new image series the fibers could be misplaced, and similarly if the fibers are visualized in the MPR on B, the positioning will not be correct. The misalignment will be equal to the rotations/translations necessary to align A with B.
The firm distributed the recall notification letter by email. The letter identified the affected product, problem and the actions to be taken. For questions contact NordicNeurolab Customer Service or if you would like to arrange a return [email protected].
Worldwide Distribution - US Nationwide - Saudi Arabia, Brazil, Sweden, Spain, Russia, Belgium, Norway, South Korea, Japan
36 licenses