FDA Recall Open, Classified

Medline convenience kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN. labeled as: IV START BOX, REF ACC010665

Recall: Z-2044-2025 · Initiated May 23, 2025

Recall

Recall Number
Z-2044-2025
Event Number
97027
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
LRS
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 23, 2025
Posted
June 30, 2025
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

Medline convenience kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN. labeled as: IV START BOX, REF ACC010665

Reason

Medline Industries, LP. is issuing a recall for specific Medline kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN which was identified that during insertion of the Catheter, the needle is slow to retract or fails to retract.

Action

Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 05/23/2025 via email and USPS mail. The notice explained the issue, risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product.2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form.Website link: https://recalls.medline.comRecall Reference #: R-25-106-FGX2Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable.4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]."

Distribution

US Nationwide distribution in the states of IN, KY, CA, MI, TX, WA.

Quantity

36 units