SELECTSILICONE" 100% SILICONE FOLEY CATHETER 2-WAY 8 Fr 3 ml - Product Usage: Medline Silicone Foley catheter is a flexible tubular device that is inserted through the urethra and used to drain urine into a collection device.

Recall: Z-2043-2020 · Initiated May 28, 2019

Recall

Recall Number
Z-2043-2020
Event Number
84791
Firm
Medline Industries Inc
FEI Number
1417592
Product Code
EZL
Status
Terminated
Root Cause
Manufacturing material removal
Initiated
May 28, 2019
Terminated
January 30, 2023
Address
Three Lakes Drive, Northfield, IL, 60093

Description

SELECTSILICONE" 100% SILICONE FOLEY CATHETER 2-WAY 8 Fr 3 ml - Product Usage: Medline Silicone Foley catheter is a flexible tubular device that is inserted through the urethra and used to drain urine into a collection device.

Reason

Loose silicone particulate was found to be present on the shaft of the silicone catheters.

Action

Medline Industries, Inc. sent formal letters with response forms to each consignee affected via 1st class mail on May 28, 2019. Each consignee was also provided with a response form to return via fax or email. The recalled product that is will be quarantined at the corporate office located at Three Lakes Drive, Northfield, IL 60093-2753 pending destruction.

Distribution

US Nationwide distribution including in the states of CO, IL, KY, MO, NJ, OR, TN, TX, VA, WA.