FDA Recall Open, Classified

MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).

Recall: Z-2040-2024 · Initiated April 15, 2024

Recall

Recall Number
Z-2040-2024
Event Number
94557
Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 4, Hayozma St Tirat Carmel Israel
FEI Number
3004571294
Product Code
KPS
Status
Open, Classified
Root Cause
Device Design
Initiated
April 15, 2024
Posted
June 7, 2024

Description

MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).

Reason

There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.

Action

The recalling firm issued letters on and dated 4/15/2024 via FedEx informing the consignee of the safety issue and actions to be taken by the customer/user. The actions state the consignee can continue to use the system but until GE HealthCare implements the correction for this potential issue on the affected device and before conducting any servicing activities that involve movement of the covers, GE HealthCare Service should be contacted for guidance. An acknowledgement form was enclosed for completion and return. The letter contained an embedded photograph of the device showing where to find the manufacturing date to determine if it is an affected device. The letter also informs the consignee they will be contacted by a GE HealthCare representative to arrange for the correction.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, NC, ND, NV, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Belgium, Brazil, Canada, Czech Republic, Finland, France, India, Israel, Italy, Japan, Malaysia, Malta, Poland, Singapore, Spain, Taiwan, Turkey, and United Kingdom.

Quantity

133 devices