SYS IVK, Bio mCT-X 3R->4R Upgrade, MATERIAL NUMBER 10250745 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Recall
- Recall Number
- Z-2036-2016
- Event Number
- 74192
- Firm
- Siemens Medical Solutions USA Inc.
- FEI Number
- 1034973
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- April 22, 2016
- Posted
- June 21, 2016
- Terminated
- May 8, 2017
- Address
- 810 Innovation Dr, Knoxville, TN, 37932-2562
Description
SYS IVK, Bio mCT-X 3R->4R Upgrade, MATERIAL NUMBER 10250745 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
Siemens sent a Customer Safety Advisory Letter on April 22, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Siemens representative at the contact numbers provided in the letter.
Worldwide Distribution - US Nationwide
2 units