FDA Recall Terminated

Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.

Recall: Z-2036-2014 · Initiated June 30, 2014

Recall

Recall Number
Z-2036-2014
Event Number
68696
Firm
Steris Corporation
FEI Number
1527821
Product Code
MEC
Status
Terminated
Root Cause
Device Design
Initiated
June 30, 2014
Posted
July 10, 2014
Terminated
July 1, 2015
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.

Reason

STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.

Action

Customers will be notified via onsite visits by STERIS Field Service Technicians to each affected Customer site. These visits will be documented and monitored until completion.

Distribution

US Nationwide Distribution - in the states of AL, AR, CA, DE, FL, GA, IL, IA, LA, MA, MI, NJ, NY, OH, OK, OR, SC, TX, VA. and in the country of Canada.

Quantity

37