FDA Recall Terminated

Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan

Recall: Z-2034-2017 · Initiated March 23, 2017

Recall

Recall Number
Z-2034-2017
Event Number
76828
Firm
Sysmex America Inc
FEI Number
3009711478
Product Code
GKZ
Status
Terminated
Root Cause
Component design/selection
Initiated
March 23, 2017
Posted
April 14, 2017
Terminated
August 22, 2019
Address
577 Aptakisic Rd, Lincolnshire, IL, 60069-4325

Description

Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan

Reason

Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.

Action

Sysmex sent an Field Correction letter dated April 2017. Product will not be returned, but corrected at the customer site. The correction requires a service visit that will be documented in the Sysmex service Customer Relationship Management (CRM) software. The service report is accepted by the customer at the time of service. Completion of the service records will be tracked by reports generated from the Service CRM. For further Technical Assistance or questions please call (1 888) 879-7639.

Distribution

Nationwide, Canada

Quantity

3,682 units