Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
Recall
- Recall Number
- Z-2034-2017
- Event Number
- 76828
- Firm
- Sysmex America Inc
- FEI Number
- 3009711478
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- March 23, 2017
- Posted
- April 14, 2017
- Terminated
- August 22, 2019
- Address
- 577 Aptakisic Rd, Lincolnshire, IL, 60069-4325
Description
Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.
Sysmex sent an Field Correction letter dated April 2017. Product will not be returned, but corrected at the customer site. The correction requires a service visit that will be documented in the Sysmex service Customer Relationship Management (CRM) software. The service report is accepted by the customer at the time of service. Completion of the service records will be tracked by reports generated from the Service CRM. For further Technical Assistance or questions please call (1 888) 879-7639.
Nationwide, Canada
3,682 units