ADVANTAGE Turbo 2-Button Shaver. Rx Only, Made In USA, NON STERILE. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908 USA.

Recall: Z-2026-2010 · Initiated October 29, 2009

Recall

Recall Number
Z-2026-2010
Event Number
54521
Firm
Linvatec Corp. dba ConMed Linvatec
FEI Number
1017294
Product Code
GEY
Status
Terminated
Root Cause
Other
Initiated
October 29, 2009
Posted
July 17, 2010
Terminated
August 19, 2010
Address
11311 Concept Blvd, Largo, FL, 33773-4908

Description

ADVANTAGE Turbo 2-Button Shaver. Rx Only, Made In USA, NON STERILE. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908 USA.

Reason

ConMed Linvatec is recalling Advantage, 2-Button Turbo Handpiece Catalog Number D9924. These products may self-activate due to moisture intrusion into the housing body. This issue has been isolated to products manufactured between June 7, 2004 and June 1, 2006.

Action

ConMed Linvatech notified affected customers by letter sent by FedEx next day delivery. Customers were asked to complete a response form and indicated if they have the product or have transferred the product. A Service Request number will be assigned for the returned product. Additional questions are directed to the firm's Customer Service department at 800-535-8536.

Distribution

Worldwide distribution. Products were distributed domestically throughout the U.S. Also United Arab Emirates, Argentina, Australia, Belgium, Switzerland, Chile, China, Germany, Denmark, Spain, France, United Kingdom, Hong Kong, India, Italy, Japan, Lithuania, Morocco, Netherlands, New Zealand, Singapore, Turkey, South Africa.

Quantity

2,373