FDA Recall Terminated

Icelock Expulsion Valve 551 Model L-551000 The valve has a push button for suction release to release a lower limb prosthesis.

Recall: Z-2026-2009 · Initiated June 23, 2009

Recall

Recall Number
Z-2026-2009
Event Number
52773
Firm
Ossur Engineering, Inc
FEI Number
3002440200
Product Code
ISP
Status
Terminated
Root Cause
Process change control
Initiated
June 23, 2009
Posted
September 4, 2009
Terminated
March 6, 2012
Address
910 Burstein Dr, Albion, MI, 49224-4011

Description

Icelock Expulsion Valve 551 Model L-551000 The valve has a push button for suction release to release a lower limb prosthesis.

Reason

The plastic caps have come off or become loose when the user stepped into their prosthesis.

Action

Customers were sent a Safety Notice Voluntary Product Recall, dated June 23, 2009, that advised them to return any unused valves to Ossur for a free replacement. Those using valves in combination with a cushion liner and suspension sleeve should replace immediately if problems are experienced, otherwise replace at next routine service. Customers using valve in combination with an Iceross Seal-In liner should replace ASAP and return used valves to Ossur for free replacement. Customers with questions can call Darlyne Roa 1-800-233-6263 ext 1023743.

Distribution

Worldwide Distribution

Quantity

4,647