Aquatec Fixed Backrest - accessory for Mobile Shower and Toilet commode Aquatec Ocean, Ocean VIP and Ocean E-VIP family.
Recall: Z-2023-2014
·
Initiated June 25, 2014
Recall
- Recall Number
- Z-2023-2014
- Event Number
- 68399
- Firm
- Invacare Corporation
- FEI Number
- 1525712
- Product Code
- INN
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 25, 2014
- Posted
- July 9, 2014
- Terminated
- April 15, 2015
- Address
- 1200 Taylor St, Elyria, OH, 44035-6248
Description
Aquatec Fixed Backrest - accessory for Mobile Shower and Toilet commode Aquatec Ocean, Ocean VIP and Ocean E-VIP family.
Reason
The Aquatec Fixed Backrest withstands a load of only 250 newtons at uneven load. This may result in the Fixed Backrest becoming detached from the frame.
Action
The recall notification will be sent to the two consignees informing them of the recall and to notify their customers. The letter will include a provider letter, customer letter, provider response card and impacted units listing.
Distribution
Distributed in PA and Canada.
Quantity
465 units