Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany 95478. The 550 TxT Treatment Table is to be used as an integral part of the Siemens linear accelerator systems to position the patient accurately according to the treatment plan for various radiation therapy treatment techniques.
Recall
- Recall Number
- Z-2023-2011
- Event Number
- 55577
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2937457
- Product Code
- JAI
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- March 31, 2010
- Posted
- April 19, 2011
- Terminated
- March 1, 2012
- Address
- 4040 Nelson Ave, Concord, CA, 94520-1200
Description
Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany 95478. The 550 TxT Treatment Table is to be used as an integral part of the Siemens linear accelerator systems to position the patient accurately according to the treatment plan for various radiation therapy treatment techniques.
Potential for causing overheating of the power supply during treatment, and potential for unexpected table descent.
Siemens issued a Customer Information letter dated March 21, 2010. There were two safety issues. There were two occurrences of smoke caused by an overheating of the power supply during patient treatment. Also, there had been a report of an unexpected lowering of the 550TxT table in patient rescue mode. With the released kit that accompanies this letter, Siemens has implemented a minor modification on the 550TxT to include an additional fuse within the power supply circuit to address the first issue described above. They also included an additional modification to address the lowering of the table by the integration of additional relays. As soon as the "move enable" button is released, the table brake will be enabled which is controlled by the implemented relays. No further action is required for the continued operations of the 550TxT table. Further questions regarding this recall please call (925) 246-8407
Worldwide Distribution -- Including USA, states of NJ, WI, NY, and MA.
326 units