FDA Recall Terminated

Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany 95478. The 550 TxT Treatment Table is to be used as an integral part of the Siemens linear accelerator systems to position the patient accurately according to the treatment plan for various radiation therapy treatment techniques.

Recall: Z-2023-2011 · Initiated March 31, 2010

Recall

Recall Number
Z-2023-2011
Event Number
55577
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2937457
Product Code
JAI
Status
Terminated
Root Cause
Component design/selection
Initiated
March 31, 2010
Posted
April 19, 2011
Terminated
March 1, 2012
Address
4040 Nelson Ave, Concord, CA, 94520-1200

Description

Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany 95478. The 550 TxT Treatment Table is to be used as an integral part of the Siemens linear accelerator systems to position the patient accurately according to the treatment plan for various radiation therapy treatment techniques.

Reason

Potential for causing overheating of the power supply during treatment, and potential for unexpected table descent.

Action

Siemens issued a Customer Information letter dated March 21, 2010. There were two safety issues. There were two occurrences of smoke caused by an overheating of the power supply during patient treatment. Also, there had been a report of an unexpected lowering of the 550TxT table in patient rescue mode. With the released kit that accompanies this letter, Siemens has implemented a minor modification on the 550TxT to include an additional fuse within the power supply circuit to address the first issue described above. They also included an additional modification to address the lowering of the table by the integration of additional relays. As soon as the "move enable" button is released, the table brake will be enabled which is controlled by the implemented relays. No further action is required for the continued operations of the 550TxT table. Further questions regarding this recall please call (925) 246-8407

Distribution

Worldwide Distribution -- Including USA, states of NJ, WI, NY, and MA.

Quantity

326 units