FDA Recall Terminated

LMA Esophageal Intubation Detector (EID), Catalog Number EID100 The device is used to assist verification of placement of the endotracheal tube or esophageal/tracheal double lumen tube. This device is to be used as an adjunct to assess intubation. Its purpose is not to eliminate clinical judgment

Recall: Z-2022-2012 · Initiated May 23, 2012

Recall

Recall Number
Z-2022-2012
Event Number
62337
Firm
LMA North America Inc
FEI Number
3001338204
Product Code
BTR
Status
Terminated
Root Cause
Process control
Initiated
May 23, 2012
Posted
July 17, 2012
Terminated
December 26, 2012
Address
4660 La Jolla Village Dr, Ste 900, San Diego, CA, 92122-4608

Description

LMA Esophageal Intubation Detector (EID), Catalog Number EID100 The device is used to assist verification of placement of the endotracheal tube or esophageal/tracheal double lumen tube. This device is to be used as an adjunct to assess intubation. Its purpose is not to eliminate clinical judgment

Reason

The recall was initiated because LMANA has received a complaint that the Esophageal Intubation Detector (EID100) has a potential to create a false positive if used on a patient during confirmation of intubation.

Action

LMA North America Inc. sent a recall letter dated May 23, 2012, for US customers, and May 24, 2012, for International customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to fill out the bottom portion of the recall letter and promptly fax it back to LMA North America at (858) 622-4130. If customers have units of EID100 from the affected lots remaining in stock, customers were instructed to arrange for immediate return by contacting LMANA's customer service department at (800) 788-7999, menu selection #2. For further questions please contact Customer Service Representative at (800) 788-7999.

Distribution

Worldwide Distribution--USA (nationwide) and the countries of Canada, United Kingdom, ZA, and Australia.

Quantity

18,799 units