FDA Recall
Terminated
Intuitive Surgical Inc. Large Clip Applier instruments used in conjunction with da Vinci S and da Vinci SI Surgical Systems Intuitive Surgical Inc., 'Sunnyvale, CA 94086
Recall: Z-2020-2014
·
Initiated June 13, 2014
Recall
- Recall Number
- Z-2020-2014
- Event Number
- 68553
- Firm
- Intuitive Surgical, Inc.
- FEI Number
- 3001675293
- Product Code
- NAY
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- June 13, 2014
- Posted
- July 8, 2014
- Terminated
- December 1, 2014
- Address
- 1266 Kifer Rd, Bldg 100, Sunnyvale, CA, 94086-5304
Description
Intuitive Surgical Inc. Large Clip Applier instruments used in conjunction with da Vinci S and da Vinci SI Surgical Systems Intuitive Surgical Inc., 'Sunnyvale, CA 94086
Reason
Manufacturing variation in the grip assembly of the Large Clip applier which can lead to the instrument jamming during insertion when used with the 8 mm cannula or with certain reducer accessories..
Action
Urgent Medical Device Correction letters were sent on June 18, 2014 to all affected sites via trackable mail.
Distribution
Worldwide Distribution.
Quantity
5290 for part 420230-04; 1964 for part 420240-06