FDA Recall Open, Classified

Servo-n Ventilator System. Model Number: 6694800.

Recall: Z-2014-2025 · Initiated May 30, 2025

Recall

Recall Number
Z-2014-2025
Event Number
97032
Firm
Maquet Critical Care AB Rontgenvagen
FEI Number
3002808128
Product Code
CBK
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 30, 2025
Posted
June 27, 2025
Address
2 Solna Sweden

Description

Servo-n Ventilator System. Model Number: 6694800.

Reason

Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.

Action

Maquet Critical Care AB, a subsidiary of Getinge, notified consignees on about 05/30/2025 via FedEx letter. Consignees were instructed to configure all affected Servo ventilators with "Extended leakage test" set to "Off", per chapter 8.2.4 Configuration in the User's manuals and all users are recommended to follow the updated instructions for use of the User's manuals chapter 9.5.1 Patient circuit test, which is provided. Consignees are instructed to ensure all devices in inventory have been properly addressed as described above, to ensure the message is forwarded to all affected personnel and to individuals if units were transferred. Additionally, consignees were requested to complete and return the provided response form.

Distribution

US Nationwide distribution and Puerto Rico. No international distribution.

Quantity

7,688 units