FDA Recall Terminated

Abbott Molecular Bar Code Scanner User's Guide; instructions for setting up the Bar Code Scanner LN 6L89-01, a component of the Abbott m2000 Instrument System; Abbott Molecular Inc., Des Plaines, IL 60018 USA; list number 06L88-02. Language configuration bar codes - Intended Use: The bar code scanner is a hand-held device that provides a convenient means of scanning sample bar codes to allow positive sample identification and is supplementary to the computer keyboard. Both the scanner and the keyboard transmit data to a System Control Center (SCC) through a single wedge (cable with branched connectors). The bar code scanner must be configured to match the keyboard language being used. In its default configuration set by the manufacturer (Jadak), the scanner is in English format. If a keyboard other than the English keyboard is being used, the user is required to use the scanner to scan an appropriate language configuration bar code provided in the User's Guide to match the language of the keyboard. Users in the U.S have the ability to convert to another language.

Recall: Z-2013-2012 · Initiated January 31, 2012

Recall

Recall Number
Z-2013-2012
Event Number
62329
Firm
Abbott Molecular
FEI Number
3005248192
Product Code
JQW
Status
Terminated
Root Cause
Error in labeling
Initiated
January 31, 2012
Posted
July 14, 2012
Terminated
December 4, 2012
Address
1300 E Touhy Ave, Des Plaines, IL, 60018-3315

Description

Abbott Molecular Bar Code Scanner User's Guide; instructions for setting up the Bar Code Scanner LN 6L89-01, a component of the Abbott m2000 Instrument System; Abbott Molecular Inc., Des Plaines, IL 60018 USA; list number 06L88-02. Language configuration bar codes - Intended Use: The bar code scanner is a hand-held device that provides a convenient means of scanning sample bar codes to allow positive sample identification and is supplementary to the computer keyboard. Both the scanner and the keyboard transmit data to a System Control Center (SCC) through a single wedge (cable with branched connectors). The bar code scanner must be configured to match the keyboard language being used. In its default configuration set by the manufacturer (Jadak), the scanner is in English format. If a keyboard other than the English keyboard is being used, the user is required to use the scanner to scan an appropriate language configuration bar code provided in the User's Guide to match the language of the keyboard. Users in the U.S have the ability to convert to another language.

Reason

In the Abbott Molecular Bar Code Scanner User's Guide, List Number 6L88-02/09, the section Modifying the Keyboard Format for Bar Code Scanner LN 6L89-01 contains incorrect bar codes that can prevent the correct configuration of the bar code scanner language format. The Abbott m2000 System is intended for use in performing nucleic acid testing in clinical laboratories. It is comprised of the Ab

Action

Abbott Molecular Inc. sent an Urgent Field Safety Notice letter dated January 31, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed to replace the affected pages of the Bar Code Scanner User's Guide with the replacement pages provided with the letter, and immediately perform the Modifying the Keyboard Format section with these replacement pages. The accounts were requested to complete the enclosed reply form and fax it to (847) 775-6728 to acknowledge receipt, understanding and implementation of the information provided in the notice.

Distribution

California, Florida, Maryland, Oregon and Virginia, and internationally to Belgium, Brazil, China, France, Germany, Hong Kong, Japan, Malaysia, Mexico, Qatar, Switzerland and United Kingdom

Quantity

100 units