FDA Recall Terminated

Microstream FILTERLINE SET ADU/PED 25UN Part Number: XS04620-01 Oridion XS04661 Physio Control

Recall: Z-2012-2010 · Initiated May 26, 2010

Recall

Recall Number
Z-2012-2010
Event Number
55790
Firm
Oridion Medical
FEI Number
3001227743
Product Code
cck
Status
Terminated
Root Cause
Process design
Initiated
May 26, 2010
Posted
July 13, 2010
Terminated
June 21, 2012
Address
1987 Ltd. 7 Hamarpe St. P.O. Box 45025, Har Hotzvim Industrial Park Jerusalem Israel

Description

Microstream FILTERLINE SET ADU/PED 25UN Part Number: XS04620-01 Oridion XS04661 Physio Control

Reason

The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment.

Action

Oridion notified customers by letter on 5/26/10 and requested that they remove the FilterLines from the related production lots from the field and an immediate replacement with FilterLines from subsequent production lots. Questions are directed to the responsible person at the supplier named in the letter, or to Oridion Capnography at 888-674-3466.

Distribution

AZ, CA, FL, HI, LA, , MA, MO, NJ, OH, PA, WA, WI,

Quantity

142,875 units (5715 boxes/25 )