FDA Recall Terminated

BrightView XCT , Model Number 882482

Recall: Z-2011-2019 · Initiated June 10, 2019

Recall

Recall Number
Z-2011-2019
Event Number
83201
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
KPS
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
June 10, 2019
Terminated
July 24, 2020
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

BrightView XCT , Model Number 882482

Reason

An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.

Action

Urgent Field Safety Notice Medical Device Correction letters dated 6/4/19 were sent to customers. Discontinue Use of the BrightView Handcontroller. Remove the handcontroller from the charging station and store in a secure location inaccessible by technicians, at least 20 meters (65 feet) away from the system and outside the imaging room. If 20 meters (65 feet) is not possible, then store the handcontroller in an area that is shielded, such as a hot lab. Do not dispose of the hand controller, as it will be needed once a solution is identified. Instead of the physical handcontroller, use the virtual handcontroller located on the touchscreen monitor attached to the gantry as shown in Figure 3. Philips Healthcare is distributing this FSN to all affected customers/users and will deploy a technical solution addressing the issue upon completion of the investigation. If you need any further information or support concerning this issue, please contact your local Philips representative: For SUPPORT in North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts).

Distribution

Worldwide Distribution - US Nationwide

Quantity

427