FDA Recall Terminated

BIOGRAPH 40 TruePoint w/TrueV, Material Number 10097304 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Recall: Z-2009-2016 · Initiated April 22, 2016

Recall

Recall Number
Z-2009-2016
Event Number
74192
Firm
Siemens Medical Solutions USA Inc.
FEI Number
1034973
Product Code
KPS
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 22, 2016
Posted
June 21, 2016
Terminated
May 8, 2017
Address
810 Innovation Dr, Knoxville, TN, 37932-2562

Description

BIOGRAPH 40 TruePoint w/TrueV, Material Number 10097304 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Reason

Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.

Action

Siemens sent a Customer Safety Advisory Letter on April 22, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Siemens representative at the contact numbers provided in the letter.

Distribution

Worldwide Distribution - US Nationwide

Quantity

46 units