Cobham AMOGS Advanced Medical Oxygen Generating System, 70309-03900-101. Concentrator, Oxygen, Molecular Sieve, S/A P/N: 70309-03900-101, Mfg P/N 3261165-0101. The firm name on the label is Carleton Life Support Systems Inc., Davenport, IA.
Recall
- Recall Number
- Z-2009-2009
- Event Number
- 52766
- Firm
- Carleton Life Support Systems Inc
- FEI Number
- 3002840531
- Product Code
- CAW
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 24, 2009
- Posted
- September 16, 2009
- Terminated
- March 22, 2012
- Address
- 2734 Hickory Grove Rd, Davenport, IA, 52804-1203
Description
Cobham AMOGS Advanced Medical Oxygen Generating System, 70309-03900-101. Concentrator, Oxygen, Molecular Sieve, S/A P/N: 70309-03900-101, Mfg P/N 3261165-0101. The firm name on the label is Carleton Life Support Systems Inc., Davenport, IA.
The backup oxygen supply control valve on the oxygen concentrator could fail resulting in it sticking in the open or closed positions.
Cobham Life Support Systems, Inc. issued letters to their two customers via e-mail. One customer letter was dated July 24, 2009 and the other was dated July 28, 2009. Both letters notified the customer of the reason for the recall and that the usage of the concentrator must be restricted to exclude actual use with a patient until corrected. The letter informs customers they are redesigning the backup supply control valve and will coordinate the retrofit in the future. For further information, contact Cobham Life Support at 1-563-383-6421.
Nationwide Distribution -- CT and AL.
23 units