FDA Recall Open, Classified

NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed

Recall: Z-2007-2023 · Initiated May 19, 2023

Recall

Recall Number
Z-2007-2023
Event Number
92300
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
FNL
Status
Open, Classified
Root Cause
Device Design
Initiated
May 19, 2023
Posted
June 26, 2023
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed

Reason

Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.

Action

Baxter issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/19/2023 by letter (USPS). The notice explained the issue, hazard, and requested the following: Baxter is asking that you take the following actions: 1. Share this letter with your IT department (or other relevant personnel) to determine if you have the impacted software version and contact Baxter Technical Support at 1-800-445-3720, available twenty-four hours a day, seven days a week, to schedule a software upgrade which addresses the identified issue. 2. You may continue to use your Voalte Patient Safety Software (VPS) (with your Centrella bed) until a software upgrade is completed by implementing either of these options into your clinical bedside workflow: Option 1: When working with a patient in the room, and a bed exit alarm is activated, only silence the bed, and extend the silence timer instead of turning off bed exit. " At the Graphical Room Station (GRS), click Menu " Click Manage Bed " On the Manage Bed screen, click the Safety Alerts ON / OFF slider button and the Delay button will pop up. Extend the silence period for the desired amount of time. Option 2: After working with a patient in the room, and prior to exiting the room, enable Patient Safety at the Graphical Room Station (GRS). " At the Graphical Room Station, click Menu " Click Manage Bed on the Manage Bed screen, toggle the Safety Alerts button to ON . For general questions regarding this communication, contact Baxter Technical Support at 1-800-445-3720, available twenty-four hours a day, seven days a week.

Distribution

US Nationwide distribution in the states of OR, SC, VA, TN, GA, NC, CO, NC, FL.

Quantity

7 software versions