FDA Recall Terminated

MC5057 12ft/3.6 meters Universal Cable. Made in USA. NON STERILE Rx Only. ConMed Linvatec, 11311 Concept Blvd, Largo, FL 33773. The MC5057 Universal Cable, 12 ft, is used to connect electric handpieces to the power source. The cable is designed with a button feature that allows the user to increase or decrease the operating speed of the handpiece in use. In select handpieces, the cord button feature can select the function, i.e. screw, tap, or drill mode.

Recall: Z-2007-2009 · Initiated July 31, 2009

Recall

Recall Number
Z-2007-2009
Event Number
52812
Firm
Linvatec Corp. dba ConMed Linvatec
FEI Number
1017294
Product Code
HWE
Status
Terminated
Root Cause
Component design/selection
Initiated
July 31, 2009
Posted
October 1, 2009
Terminated
July 11, 2011
Address
11311 Concept Blvd, Largo, FL, 33773-4908

Description

MC5057 12ft/3.6 meters Universal Cable. Made in USA. NON STERILE Rx Only. ConMed Linvatec, 11311 Concept Blvd, Largo, FL 33773. The MC5057 Universal Cable, 12 ft, is used to connect electric handpieces to the power source. The cable is designed with a button feature that allows the user to increase or decrease the operating speed of the handpiece in use. In select handpieces, the cord button feature can select the function, i.e. screw, tap, or drill mode.

Reason

Sufficiently worn/damaged Universal Cable with product number MC5057, when used with various ConMed Linvatec electric Handpieces, may cause the handpiece to self activate.

Action

ConMed Linvatec notified customers via letter, cautioning the users to inspect cords for signs of excessive wear or damage and urging them to discontinue use and return cables when appropriate. Direct questions to Linvatec Corporation by calling 1-800-237-0169.

Distribution

Worldwide Distribution including United States.

Quantity

12,205 units