Focal SIM Planning of radiation therapy
Recall
- Recall Number
- Z-2002-2012
- Event Number
- 62433
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- MUJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 21, 2012
- Posted
- July 12, 2012
- Terminated
- April 1, 2014
- Address
- 4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, GA, 30092-3011
Description
Focal SIM Planning of radiation therapy
Incorrect patient shift directions when the Setup reference dialog is printed out when the DICOM coordinates option is enabled for reports.
IMPAC Medical Systems, Inc. sent an Important Safety Notice LINFFM0001/1.0 on June 21, 2012, to consignees on concerning a patch to the software. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to not use the DICOM coordinates option for Reports when printing the Setup Reference Diaglog. Customers were instructed go to www.elekta.com, select the SupportPlus Login at the tope of the page, enter their portal, select Downloads/Updates to download the latest patches. For questions regarding this recall call 770-670-2548.
Nationwide Distribution including CA, CT, FL, IL, IN, KY, MI, MN, MO, NJ, NY, NC, OH, OK, OR, PA, WA, and WI.
38 units