FDA Recall Open, Classified

i-STAT EG6+ cartridge; List Number: 03P77-25;

Recall: Z-2000-2026 · Initiated April 1, 2026

Recall

Recall Number
Z-2000-2026
Event Number
98635
Firm
Abbott Point Of Care Inc.
FEI Number
2245578
Product Code
CHL
Status
Open, Classified
Root Cause
Process change control
Initiated
April 1, 2026
Posted
April 29, 2026
Address
400 College Rd E, Princeton, NJ, 08540-6607

Description

i-STAT EG6+ cartridge; List Number: 03P77-25;

Reason

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

Action

On April 1, 2026, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. This Urgent Medical Device Corrective is to customers in the U.S. with the following recommended actions: " Discontinue use of the affected i-STAT EG7+, EG6+, and G3+ cartridge lots for reporting PCO and pH results. " Use an alternate cartridge type, alternate lot, or testing method. " Complete and return the Business Reply Card included with this notification to confirm receipt and understanding. If you have forwarded any of the specified i-STAT EG7+, EG6+ or G3+ cartridges to another facility, we request that you please provide a copy of this letter to them.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.

Quantity

30775 units (25625 US, 5150 OUS)