BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10528956 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Recall
- Recall Number
- Z-1999-2016
- Event Number
- 74239
- Firm
- Siemens Medical Solutions USA Inc.
- FEI Number
- 1034973
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- April 25, 2016
- Posted
- June 15, 2016
- Terminated
- February 1, 2017
- Address
- 810 Innovation Dr, Knoxville, TN, 37932-2562
Description
BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10528956 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
Siemens sent a Customer Safety Advisory Letter to all affected customers. The letter identified the affected product, problem and actions to be taken. Siemens service organization will be contacting customers to schedule the update as per the attached Update Instruction and will be provided the replacement bracket to resolve the issue. For questions contact your Siemens representative.
Worldwide Distribution - US Nationwide and the countries of China and Malaysia.
1 unit