FDA Recall Terminated

BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10528956 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Recall: Z-1999-2016 · Initiated April 25, 2016

Recall

Recall Number
Z-1999-2016
Event Number
74239
Firm
Siemens Medical Solutions USA Inc.
FEI Number
1034973
Product Code
KPS
Status
Terminated
Root Cause
Process design
Initiated
April 25, 2016
Posted
June 15, 2016
Terminated
February 1, 2017
Address
810 Innovation Dr, Knoxville, TN, 37932-2562

Description

BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10528956 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Reason

The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.

Action

Siemens sent a Customer Safety Advisory Letter to all affected customers. The letter identified the affected product, problem and actions to be taken. Siemens service organization will be contacting customers to schedule the update as per the attached Update Instruction and will be provided the replacement bracket to resolve the issue. For questions contact your Siemens representative.

Distribution

Worldwide Distribution - US Nationwide and the countries of China and Malaysia.

Quantity

1 unit