FDA Recall Terminated

300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. Product is a disposable product used with Microlab STAR series Instruments, which are microtiter diluting and dispensing devices.

Recall: Z-1999-2009 · Initiated April 9, 2009

Recall

Recall Number
Z-1999-2009
Event Number
52667
Firm
Hamilton Co
FEI Number
2915796
Product Code
JTC
Status
Terminated
Root Cause
Process control
Initiated
April 9, 2009
Posted
September 9, 2009
Terminated
January 4, 2010
Address
4970 Energy Way, Reno, NV, 89502-4123

Description

300 Microliter CO-RE Tips for use with Microlab STAR, manufactured by Hamilton Company, Reno, NV. Product is a disposable product used with Microlab STAR series Instruments, which are microtiter diluting and dispensing devices.

Reason

Product Defect-- some tips from the mold cavity 28 are too short and do not always aspirate or dispense liquid correctly.

Action

The firm initiated the recall on April 10, 2009, and issued notification to consignees via certified mail and e-mail. Customers were instructed to return the affected product to the Hamilton Company. A fax-back form was supplied with each notification. Direct questions about the recall to the Hamilton Company by calling 1-775-858-3000, extension 236.

Distribution

Worldwide Distribution -- US (states of DE, NJ, PA, MA and CA), New Zealand and Japan.

Quantity

20 cases of Part 235950 and 27 cases of Part 235902.