FDA Recall Open, Classified

MEGADYNE MEGA SOFT Pediatric Patient Return Electrode. used during electrosurgery

Recall: Z-1997-2023 · Initiated June 1, 2023

Recall

Recall Number
Z-1997-2023
Event Number
92409
Firm
Megadyne Medical Products, Inc.
FEI Number
1721194
Product Code
GEI
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 1, 2023
Posted
June 30, 2023
Address
4545 Creek Rd, Blue Ash, OH, 45242-2803

Description

MEGADYNE MEGA SOFT Pediatric Patient Return Electrode. used during electrosurgery

Reason

Firm has received reports of patient burns in surgical procedures where device was used.

Action

The firm distributed a recall notification by mail starting 06/01/2023. Customers are to provide the notice to all users of MEGADYNE MEGA SOFT and MEGA 2000 Reusable Patient Return Electrodes that participate in the cleaning, setup, and operation of affected devices. Megadyne has been made aware of 63 reports of serious patient burns since April 2018. The firm has attributed the incidence of burns to some instances that the Mega Soft Pad Instructions For Use were not being properly followed. Devices must be thoroughly rinsed after cleaning to ensure any cleaning residue is removed. Customers are asked to distribute the recall notification to all users, confirm that personnel are following the Instructions For Use, and post the provided informational attachments near the OR for staff member reference. If devices have been further distributed customers are to share the provided recall notice. Completed Business Reply Forms are to be faxed to Sedgwich at 888-214-7430 or emailed [email protected]. Any questions about the Business Reply Form can be made by phone to 888-843-0254 (Reference Event 5627). Any questions about this recall or to report a complaint can be made by phone to 1-877-384-4266, Monday thru Friday from 8:00 AM to 5:00 PM. Medical Engagement can be requested at https://www.jnjmedtech.com/mir.

Distribution

Distributed US Nationwide.

Quantity

21,100 units