Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.
Recall
- Recall Number
- Z-1997-2013
- Event Number
- 65832
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- KHE
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- July 17, 2013
- Posted
- August 22, 2013
- Terminated
- October 24, 2014
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.
Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.
Beckman Coulter sent an Urgent Product Correction letter dated July 9, 2013, to all affected customers. The letter was accompanied with a cover letter and response form, dated July 9, 2013. Customers were informed of the product under recall with product description, actions to take, contact information, the issue, impact, and resolution. Beckman Coulter conducted a risk assessment for this issue. Per the risk documentation, there is only a remote probability that use of the product will cause any adverse health consequences,and the product is not likely to lead to serious adverse health consequences. If you have any questions please call (714) 961-3777.
Worldwide Distribution - USA and the country of Canada.
68,116 units