FDA Recall Terminated

i-STAT PT/INR Cartridge Abbott Point of Care, Princeton, NJ

Recall: Z-1996-2010 · Initiated May 5, 2010

Recall

Recall Number
Z-1996-2010
Event Number
55895
Firm
Abbott Point of Care Inc.
FEI Number
2245578
Product Code
GJS
Status
Terminated
Root Cause
Other
Initiated
May 5, 2010
Posted
July 13, 2010
Terminated
September 22, 2011
Address
400 College Rd E, Princeton, NJ, 08540-6607

Description

i-STAT PT/INR Cartridge Abbott Point of Care, Princeton, NJ

Reason

A high mean bias (12-16%) was documented in 8 lots of i-STAT PT/INR cartridges. This may lead a clinician to believe that a patient is adequately anticoagulated when they are not. Inadequate anticoagluation therapy may be given and issues with clotting and/or thrombosis could occur. No reports of injury have been received.

Action

Urgent recall notifications were sent on May 13 2010. Customers were instructed to discontinue use and to return the identified lots. Additional questions are directed to Abbott Point of Care Technical Support at 800-3668020, Option 1 or your Abbott Point of Care representative.

Distribution

Worldwide distribution: Argentina, Hong Kong, China, Finland, Greece, India, Portugal, Russian Federation, Saudi Arabia, United Arab Emirates, Italy, Spain, Australia, Germany, UK, Sweden, The Netherlands, France, Canada.

Quantity

44976 in US, 3648 in rest of world.