i-STAT PT/INR Cartridge Abbott Point of Care, Princeton, NJ
Recall
- Recall Number
- Z-1996-2010
- Event Number
- 55895
- Firm
- Abbott Point of Care Inc.
- FEI Number
- 2245578
- Product Code
- GJS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 5, 2010
- Posted
- July 13, 2010
- Terminated
- September 22, 2011
- Address
- 400 College Rd E, Princeton, NJ, 08540-6607
Description
i-STAT PT/INR Cartridge Abbott Point of Care, Princeton, NJ
A high mean bias (12-16%) was documented in 8 lots of i-STAT PT/INR cartridges. This may lead a clinician to believe that a patient is adequately anticoagulated when they are not. Inadequate anticoagluation therapy may be given and issues with clotting and/or thrombosis could occur. No reports of injury have been received.
Urgent recall notifications were sent on May 13 2010. Customers were instructed to discontinue use and to return the identified lots. Additional questions are directed to Abbott Point of Care Technical Support at 800-3668020, Option 1 or your Abbott Point of Care representative.
Worldwide distribution: Argentina, Hong Kong, China, Finland, Greece, India, Portugal, Russian Federation, Saudi Arabia, United Arab Emirates, Italy, Spain, Australia, Germany, UK, Sweden, The Netherlands, France, Canada.
44976 in US, 3648 in rest of world.