HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
Recall
- Recall Number
- Z-1995-2014
- Event Number
- 68580
- Firm
- Cenorin, LLC
- FEI Number
- 3004153171
- Product Code
- LDS
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- January 24, 2013
- Posted
- July 6, 2014
- Terminated
- February 10, 2015
- Address
- 6324 S 199th Pl, Ste 107, Kent, WA, 98032-2135
Description
HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
Cenorin sent a Notification letter dated December 18, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated: Cenorin has developed an upgrade to mitigate this issue as well as provide additional process control and system monitoring benefits. Cenorin included a Return Response form with the recall letter to be completed and return. For questions contact HLD Systems Sales Representative for questions.
Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
25 HLD systems