FDA Recall Terminated

HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.

Recall: Z-1995-2014 · Initiated January 24, 2013

Recall

Recall Number
Z-1995-2014
Event Number
68580
Firm
Cenorin, LLC
FEI Number
3004153171
Product Code
LDS
Status
Terminated
Root Cause
Device Design
Initiated
January 24, 2013
Posted
July 6, 2014
Terminated
February 10, 2015
Address
6324 S 199th Pl, Ste 107, Kent, WA, 98032-2135

Description

HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.

Reason

The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.

Action

Cenorin sent a Notification letter dated December 18, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated: Cenorin has developed an upgrade to mitigate this issue as well as provide additional process control and system monitoring benefits. Cenorin included a Return Response form with the recall letter to be completed and return. For questions contact HLD Systems Sales Representative for questions.

Distribution

Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.

Quantity

25 HLD systems