Datex-Ohmeda S/5 Critical Care Monitor with L-CICU01, L-CICU02A, L-ICU05 & L-ICUO5A software. CCM Indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinavregional perfusion, B ispectral index (BIS), and neurophysiological status of all hospital patients.
Recall
- Recall Number
- Z-1994-2009
- Event Number
- 51278
- Firm
- GE Medical Systems Information Tecnology 9900 Inovation Drive RP 2122 3rd Floor Wauwatosa WI 53226
- FEI Number
- 3005860720
- Product Code
- MLD
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 6, 2006
- Posted
- September 1, 2009
- Terminated
- May 17, 2012
Description
Datex-Ohmeda S/5 Critical Care Monitor with L-CICU01, L-CICU02A, L-ICU05 & L-ICUO5A software. CCM Indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinavregional perfusion, B ispectral index (BIS), and neurophysiological status of all hospital patients.
GE Healthcare (Datex-Ohmeda) S/5 Critical Care Monitor and S/5 Compact Critical Care Monitor with L-CICU01, L-CICU02A, L-ICU05 & L-ICUO5A software had an issue where the Heart Rate (HR) was displaying as ---(3 dashes) when the monitor was used continuously for 49 days without a cold start. A missed alarm or missed HR value could cause a delay in recognition of a cardiac event that may result i
GE Healthcare Technologies issued a Marketing Bulletin "Ref: MB MS HKI 06007" dated February 6, 2006 to explain a software update.
ITALY, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRASIL, BULGARIA, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA, IRAN, ISRAEL JORDAN, LITHUANIA, ITALY, MACEDONIA, MALAYSIA, MEXICO, NETHERLAND, NICARAGUA, NORWAY, POLAND, PORTUGAL PUERTO RICO, RUSSIA, SINGAPORE, SLOVAK IA, SOUTH AFRICA, SPAIN, SWEDEN, SYRIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA.
873