FDA Recall Terminated

GE Healthcare Optima Mobile X-ray System. The GE Automatic Mobile X-Ray (AMX ) Series are intended to take exposures utilizing film or computed radiography ( CR )

Recall: Z-1993-2012 · Initiated April 4, 2012

Recall

Recall Number
Z-1993-2012
Event Number
62527
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
IZL
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
April 4, 2012
Posted
July 19, 2012
Terminated
January 10, 2013
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Optima Mobile X-ray System. The GE Automatic Mobile X-Ray (AMX ) Series are intended to take exposures utilizing film or computed radiography ( CR )

Reason

GE Healthcare has become aware of a software issue on the interface of the mobile radiographic product, Optima XR200amx, ond Optima XR220amx related to Visual indication as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-1-3.

Action

GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal and IEC standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to the user or through a GE Healthcare field engineer site visit. A GE Healthcare Service Representative will update the firmware on the system to address the issue. Your proposed corrective action plan (CAP) (Field Modification Instructions 10865) describes the rework plan that GEHC Engineers will make to the imaging system to return it to compliance with the applicable performance standard. It appears to adequately address the problem and is hereby approved.

Distribution

Worldwide Distribution (USA) nationwide

Quantity

254 units have been shipped to/installed in the U.S.