FDA Recall Open, Classified

Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C, RW82-1D, WBR82-1 and WBR82-1C]

Recall: Z-1992-2023 · Initiated May 16, 2023

Recall

Recall Number
Z-1992-2023
Event Number
92353
Firm
Draeger Medical Systems, Inc.
FEI Number
2510954
Product Code
FMT
Status
Open, Classified
Root Cause
Device Design
Initiated
May 16, 2023
Posted
June 22, 2023
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C, RW82-1D, WBR82-1 and WBR82-1C]

Reason

The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.

Action

An Urgent - Medical Device Correction dated May 2023 was sent to customers. Actions to be taken: Please ensure that all users of the above-mentioned products and other persons within your organization are made aware of this Important Safety Notice. If you have provided the products to third parties, please forward a copy of this information. Users can mitigate this risk by performing the visual check, test procedure and take appropriate actions based on the Attachment provided in this notice. Complete and return the enclosed Customer Acknowledgement and Response Form . We are continuing to evaluate this issue and will advise the affected customers when we have an alternative solution. Visual Check for Correct Mattress Placement: To ensure proper operation of the scale, please check to confirm there is a clearance between the mattress cover and the corner blocks and side panels of the Resuscitaire. 1. Ensure mattress with cover is installed in the Resuscitaire bassinet. 2. Observe edges of mattress for clearance with corner blocks and side panels, see Figure 2. 3. If there is interference, then the scale may not provide accurate weight measurements. It is recommended to use an independent scale for therapeutic decisions and treatment Test Procedure If it is unclear from the visual check if the mattress cover interferes with the corner blocks and/or side panels, impacting scale function, please perform the following test: a. Place a known calibrated weight on the center mattress (with mattress cover installed). A 5kg weight is recommended in the instructions for use. b. Execute weighing procedure and record weight. c. Repeat weighing procedure 5-10 times. d. Compare measurements. i. If measurements are within 5 grams of calibrated 5kg weight, mattress assembly is acceptable. ii. If measurements deviate more than +/- 5 grams from the calibrated 5kg weight, it is recommended to use an alternate weighing scale Contact: We regret any inconvenience this may cause.

Distribution

US Nationwide distribution including in the states of AL, AR, AZ, CA, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, PR, SD, TN, TX, VT, WA, and WI.

Quantity

936 units