Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041

Recall: Z-1992-2016 · Initiated March 31, 2016

Recall

Recall Number
Z-1992-2016
Event Number
73692
Firm
Med Tec Inc
FEI Number
1000143606
Product Code
JAI
Status
Terminated
Root Cause
Software design
Initiated
March 31, 2016
Terminated
February 3, 2017
Address
1401 8th St SE PO Box 320, Orange City, IA, 51041-7463

Description

Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041

Reason

When using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.

Action

Civco Medical Solutions sent a "Urgent Field Correction Notice" dated March 31, 2016, to all affected customers. The letter identified the product the problem and the action need to be taken by the customer. The software's other configuration, "All 6 Degrees Protura" does not have the same translational issues. The firm send notification letters on February 23, 2016. In this letter, the Product Manager verified each site's software is configured to this operational mode which operates without the identified problem. In a later and separate notification on March 31, 2016, the firm informed customers that a new software version will be installed to address the problem. No timeframe was given in the letter. For further questions, please call ( 712) 737-8688.

Distribution

Worldwide Distribution - US Distribution including the countries of Switzerland, Germany..

Quantity

9