FDA Recall Open, Classified

ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump

Recall: Z-1990-2025 · Initiated April 22, 2025

Recall

Recall Number
Z-1990-2025
Event Number
96777
Firm
ICU Medical, Inc.
FEI Number
3013319212
Product Code
FRN
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 22, 2025
Posted
June 18, 2025
Address
600 N Field Dr, Lake Forest, IL, 60045-4835

Description

ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump

Reason

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

Action

ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 05/02/2025 via hand delivery. The notice explained the issue, potential risk, and requested the following: Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification. 2. If Maximum Dose Limit is not used for a medication, the issue will not occur. If Maximum Dose Limit is used for a medication, users should take the following precautions. If you need to adjust the weight or BSA, first clear the program and then reprogram. For further inquiries: To report adverse events or product complaints: Global Complaint Management - 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or [email protected]. For any questions regarding this action: Device Correction Inquiries https://icumed.custhelp.com/app/market-action. Additional technical information or assistance: Technical Support 1-(800)-241-4002, option 3 or [email protected].

Distribution

Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

Quantity

1701 units