GE Healthcare, Proteus XR/a. Proteus XR/a Intended Use: Is intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
Recall
- Recall Number
- Z-1988-2014
- Event Number
- 68528
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- KPR
- Status
- Terminated
- Root Cause
- Equipment maintenance
- Initiated
- June 4, 2014
- Posted
- July 2, 2014
- Terminated
- May 6, 2015
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Healthcare, Proteus XR/a. Proteus XR/a Intended Use: Is intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
GE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident of a collimator fall which caused a serious patient injury.
GE Healthcare sent an Urgent Medical Device Correction letter dated June 4, 2014, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions in regard to this recall the consignees in the United States were advised to call 800-437-1171. For other countries contact your local GE Healthcare Service Representative. For questions regarding this recall call 800-437-1171.
Worldwide Distribution - USA including PR. DC, GUAM, all states except VT. Internationally to the following countries: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BENIN, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BULGARIA, BURKINA FASO, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, FRENCH POLYNESIA, GERMANY, GREECE, GUATEMALA, HONDURAS, HONG KONG, HUNGRY, ICELAND, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMHIRIYA, LITHUANIA, MALAYSIA, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UGANDA, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM YEMEN.