FDA Recall Terminated

Distributed in USA: Product Number: 40-2894: Reinforced Wound Closure, Latex Free Sterile EO Rx, 1/4" X 4 ", 6 mm X100 mm, (2 Cards/pouch, 1,000 Pouches/box); Distributed in Europe: " Product Number: 39-0058: Medicare Reinforced Wound Closure Strips 1/4" X 4 ", Sterile, Box 1000 pouches 5 strips /pouch, Pouch , 5 strips 6 mm X100 mm Sterile EO Product Number: 40-2893: Wound Closure Strips , Sterile (1,000pouches/box) Rx; Pouch Label: Wound Closure Strips 6 mm X100 mm, 1/4" X 4 ", Rx Product Usage: intended for skin closure and closure support for sutures and staples

Recall: Z-1988-2013 · Initiated July 18, 2013

Recall

Recall Number
Z-1988-2013
Event Number
65745
Firm
Aspen Surgical Products, Inc.
FEI Number
3002750084
Product Code
FPX
Status
Terminated
Root Cause
Process design
Initiated
July 18, 2013
Posted
August 20, 2013
Terminated
April 17, 2014
Address
6945 Southbelt Dr Se, Caledonia, MI, 49316-7664

Description

Distributed in USA: Product Number: 40-2894: Reinforced Wound Closure, Latex Free Sterile EO Rx, 1/4" X 4 ", 6 mm X100 mm, (2 Cards/pouch, 1,000 Pouches/box); Distributed in Europe: " Product Number: 39-0058: Medicare Reinforced Wound Closure Strips 1/4" X 4 ", Sterile, Box 1000 pouches 5 strips /pouch, Pouch , 5 strips 6 mm X100 mm Sterile EO Product Number: 40-2893: Wound Closure Strips , Sterile (1,000pouches/box) Rx; Pouch Label: Wound Closure Strips 6 mm X100 mm, 1/4" X 4 ", Rx Product Usage: intended for skin closure and closure support for sutures and staples

Reason

Aspen Surgical is recalling Wound Closures because firm could not confirm sterility, after FDA sampling indicated the product may not be sterile. Using the recalled product could lead to infection which may require further treatment including the use of antibiotics. There have been no reports of injuries or illnesses as of 7/18/2013.

Action

Aspen sent an Urgent Medical Device Recall letter dated July 18, 2013 via certified mail to all customers with directions to forward the notification and provide Aspen with sub-account information. Aspen has given the customers the option to either 1.) Return the product 2.) Destroy the product with a Certificate of Destruction to be provided to Aspen. Any returned product to Aspen will be put on MRB and dispositioned by Quality. Questions or concens should be addresses to Kelli Jona Directo of QA / RA at 888-364-7004 ext 177 [email protected] or Michelle Pulaski, Quality System and Compliance Manager 888-364-7004 ext 146 [email protected] 8-5 EST M-F.

Distribution

Worldwide Distribution - US Nationwide in the state of Florida and in the countries of , Netherlands, and Reditch Worcestershire England

Quantity

3286 boxes ( 6,353,000)