VARIANT nbs Sickle Program Resin Update CD-ROM; Model Number:250-3020, included in 250-3000; Distributed and Manufactured by Bio-Rad Laboratories, Inc. The VARIANT nbs Sickle Cell Program is intended as a qualitative screen for the presence of hemoglobins F, A, S, D, C and E in eluates of neonatal blood collected on filter paper by high-performance liquid chromatography(HPLC). The VARIANT nbs Sickle Cell Program is intended for Professional use only.
Recall
- Recall Number
- Z-1984-2011
- Event Number
- 55910
- Firm
- Bio-Rad Laboratories Inc
- FEI Number
- 2915274
- Product Code
- GKA
- Status
- Terminated
- Root Cause
- Software design (manufacturing process)
- Initiated
- April 14, 2010
- Posted
- April 14, 2011
- Terminated
- August 8, 2011
- Address
- 4000 Alfred Nobel Dr, Hercules, CA, 94547-1803
Description
VARIANT nbs Sickle Program Resin Update CD-ROM; Model Number:250-3020, included in 250-3000; Distributed and Manufactured by Bio-Rad Laboratories, Inc. The VARIANT nbs Sickle Cell Program is intended as a qualitative screen for the presence of hemoglobins F, A, S, D, C and E in eluates of neonatal blood collected on filter paper by high-performance liquid chromatography(HPLC). The VARIANT nbs Sickle Cell Program is intended for Professional use only.
The affected resin Update CD-ROMs cause all user-defined settings in the Setup/Test/Data Setup sub screen and Set/Test/Pattern Setting sub screen to be overwritten when the Update Kit procedure is performed.
Bio-Rad Laboratories sent out the Urgent Medical Device Correction notice on 4/14/2011, by Federal Express. An Update was sent on May 13th, advising customers of a minor discrepancy in the parameters included on the affected Resin Update CD-ROMs. Customers were instructed to examine your inventory for the affected reorder pack lots. Please complete and return the Medical Device Correction Response Form Questions regarding this recall please call (510) 724-6114.
Worldwide Distribution -- USA , Brazil, Canada, Caribbean, France, Italy, and United Kingdom.
658 units