FDA Recall Open, Classified

BD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use

Recall: Z-1982-2026 · Initiated April 7, 2026

Recall

Recall Number
Z-1982-2026
Event Number
98657
Firm
Becton Dickinson & Co.
FEI Number
1119779
Product Code
KZJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 7, 2026
Posted
April 29, 2026
Address
7 Loveton Cir, Sparks, MD, 21152-9212

Description

BD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use

Reason

BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.

Action

On 04/07/2026, BD sent URGENT: Medical Device Correction letters to customers and distributors. Consignees were instructed that there are no recommendations for retrospective chart review or repeat patient testing. They are to verify incubation atmospheres by incorporating the use of appropriate gas indicators as part of routine laboratory quality control practices. Each BD GasPak EZ Gas Generating Pouch System should be tested periodically for its ability to provide adequate conditions for the growth of target organisms. For CO2 incubation, laboratories may use BD CO2 Indicators (Catalog No. 271055), along with the impacted catalog 260684, to verify sufficient CO2 presence in the environment. If conditions are not confirmed or are suspected to be out of specification, laboratories should repeat incubation using alternative validated methods or established laboratory procedures to ensure organism recovery. Additionally, consignees were instructed to ensure the contents of the notification are read and understood, share this notification with the facility network and forward it to any customers to whom product may have been distributed, and complete and return the Customer Response Form. Distributors were instructed to ensure the notification is read and understood, identify all customers that purchased any affected product, and provide a copy of the attached customer letter to all customers who received the affected lots to advise them of this Product Advisory on BD's behalf. BD is investigating the root cause and will implement appropriate corrective actions to prevent recurrence of this issue. BD GasPak catalog numbers 260684 and 260685 are currently on backorder due to seal integrity challenges identified in some batches following a recent overhaul of supplier manufacturing equipment. BD anticipates resumption of product shipments by the end of March 2026, with full recovery expected by June 2026.

Distribution

Worldwide distribution - US Nationwide and the country of Hong Kong.

Quantity

25,760 units