FDA Recall Terminated

Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A) Kaluza C Perpetual License (B) Kaluza C Single Use 1 Year License (C) Kaluza C 5 User Network (D) Kaluza C 10 User Network (E) Kaluza C Perpetual Educational License (F) Kaluza C Single User 1 Year educational License (G) Kaluza C 10 User Network Educational License (H) Kaluza C 5 User Network Educational License (I) Kaluza C Flow Cytometry Software

Recall: Z-1982-2019 · Initiated September 17, 2018

Recall

Recall Number
Z-1982-2019
Event Number
82919
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JQP
Status
Terminated
Root Cause
Software Manufacturing/Software Deployment
Initiated
September 17, 2018
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following software devices: (A) Kaluza C Perpetual License (B) Kaluza C Single Use 1 Year License (C) Kaluza C 5 User Network (D) Kaluza C 10 User Network (E) Kaluza C Perpetual Educational License (F) Kaluza C Single User 1 Year educational License (G) Kaluza C 10 User Network Educational License (H) Kaluza C 5 User Network Educational License (I) Kaluza C Flow Cytometry Software

Reason

Compatibility issue between the software and Microsoft updates to Windows 7, 8, and 10. The compatibility issue causes the software to be unusable which could result in a delay of reporting results.

Action

An "Urgent Medical Device Recall" letter was sent on September 18, 2018 via email to all affected customers. Beckman Coulter ask that if you are using a LIS, please review all previous LIS transmissions which included keywords for accuracy by comparing keywords displayed in Kaluza C Information Plots to those in the LIS system. Please report any erroneous transmissions to Beckman Coulter. The recall letter also ask their customers to visit the Beckman Coulter website to download version Kaluza C 1.1 and install on all affected systems. Installation of Kaluza C version 1.1 will prevent software crashes and result in proper keyword value being transmitted for all samples 1. Navigate to http://www.kaluzasoftware.com and click the "Download" icon. 2. Click "Download Kaluza C1.1" and fill in the requested information. 3. Click the link in the registration email to download Kaluza C version 1.1. 4. Install Kaluza C version 1.1 on systems used for analysis. If you have questions regarding the notification, please contact 1-800-369-0333.

Distribution

US: CA and NY OUS: United Kingdom, Switzerland and Germany

Quantity

41 units