FDA Recall Terminated

GE Healthcare Innova IGS 530. Angiographic X-ray systems used in generating fluoroscopic and rotational images of human anatomy for diagnostic, interventional and image-guided surgical procedures.

Recall: Z-1981-2015 · Initiated May 29, 2015

Recall

Recall Number
Z-1981-2015
Event Number
71508
Firm
GE Healthcare
FEI Number
2126677
Product Code
IZI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
May 29, 2015
Posted
July 3, 2015
Terminated
October 28, 2015
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Innova IGS 530. Angiographic X-ray systems used in generating fluoroscopic and rotational images of human anatomy for diagnostic, interventional and image-guided surgical procedures.

Reason

GE Healthcare has recently become aware of a potential safety issue that may result in the loss of C-arm gantry motion capability on fluoroscopic Interventional imaging systems (Innova, Optima and Discovery).

Action

The firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter on 5/29/2015- GEHC Ref#12231- dated May 29, 2015 to their Consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and contact Information. The customers were instructed that you may continue to use the system and avoid rapid direction changes with the gantry motions and prevent potential collisions with other equipment. If the failure occurs during interventional examination, please ensure that you have established procedures for handling patients in case of the loss of gantry motion as stated in your product labeling. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for this correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, GA, IL, IN, KS, MS, NV, OH, OK, OR, PA, TN, TX, VA, WA and WI; and countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAPAN, RUSSIA, UNITED KINGDOM, and VIETNAM.

Quantity

17