FDA Recall Open, Classified

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Recall: Z-1980-2025 · Initiated May 8, 2025

Recall

Recall Number
Z-1980-2025
Event Number
96851
Firm
Ohio Medical Corporation
FEI Number
1419185
Product Code
CAW
Status
Open, Classified
Root Cause
Device Design
Initiated
May 8, 2025
Posted
June 13, 2025
Address
1111 Lakeside Dr, Gurnee, IL, 60031-2489

Description

GCE HEALTHCARE Zen-O lite, Portable Oxygen Concentrator, Model RS-00608-X-S

Reason

Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.

Action

Ohio Medical issued a Field Safety Notice to its consignees on 05/08/2025 followed by telephone calls on 05/13/2025. The notice explained the problem with the device, potential risk, and requested the devices be returned for correction. The notice further stated: "The Field Safety Notice is communicated to all impacted customers and is communicated to competent authorities in the affected markets. Receivers should pass this FSN on to all persons with a need to be advised from within your organization or to any organization where the potentially affected products have been supplied. Please consider end users and others within the supply chain in the circulation of this notice."

Distribution

US Nationwide distribution in the states of AR, FL, VA, ND, MS, MN, CA, NV, KY.

Quantity

36 units