CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.
Recall
- Recall Number
- Z-1979-2014
- Event Number
- 68459
- Firm
- Nihon Kohden America Inc
- FEI Number
- 2080783
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 3, 2014
- Posted
- July 2, 2014
- Terminated
- April 28, 2015
- Address
- 15353 Barranca Pkwy, Irvine, CA, 92618-2216
Description
CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
Nihon Kohden sent an Urgent Medical Device Safety Notification and Field Correction letter dated June 3, 2014 to affected customers. Recall notification letter, dated June 3, 2014 was hand-delivered to site by Nihon personnel. The letter provided the following information: product description with codes and software version, reason for recall, field correction information and actions to be taken, MedWatch reporting information, and contact information for recall.
Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.