FDA Recall Terminated

Advincula Delineator with Ultem Plastic Soft Cup size 3.0 cm; Part No. AD750SC-KE30 The Advincula Delineator Uterine Manipulator is a single-use disposable device designed specifically for Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH), and/or Laparoscopic Supra-Cervical Hysterectomy (LSCH).

Recall: Z-1978-2019 · Initiated June 14, 2019

Recall

Recall Number
Z-1978-2019
Event Number
83105
Firm
CooperSurgical, Inc.
FEI Number
1216677
Product Code
HEW
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
June 14, 2019
Terminated
May 19, 2020
Address
95 Corporate Dr, Trumbull, CT, 06611-1350

Description

Advincula Delineator with Ultem Plastic Soft Cup size 3.0 cm; Part No. AD750SC-KE30 The Advincula Delineator Uterine Manipulator is a single-use disposable device designed specifically for Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH), and/or Laparoscopic Supra-Cervical Hysterectomy (LSCH).

Reason

The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as lacerations.

Action

On June 14, 2019, the firm, CooperSurgical, notified customers of the recall via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were informed that the soft, flexible cup of the Advincula Delineator may crack when subjected to excessive compression during use. The failure mode could lead to serious health consequences such as lacerations. Customers are asked to do the following; - Quarantine affected products - Complete the enclosed Acknowledgement and Receipt form and return to Cooper Surgical via email: [email protected] or fax to 203.601.9870 ATTN: Product Surveillance, even if you do not have affected product. - Once the firm has received your acknowledgement form, it will arrange for product returns and replacements, if requested. If you have questions regarding this recall, you may contact the firm at 2036015200 ext. 3300.

Distribution

US distribution to state of: AZ, CA, GA, IN, LA, MA, NC, NH, NJ, NY, OH, PA, VA, and WA..

Quantity

213 units in total