FDA Recall Completed

NextSeq 550Dx Instrument, Catalog # 20005715, contaning Universal Copy Service software

Recall: Z-1976-2023 · Initiated April 5, 2023

Recall

Recall Number
Z-1976-2023
Event Number
92233
Firm
Illumina, Inc.
FEI Number
3003218906
Product Code
PFF
Status
Completed
Root Cause
Software Manufacturing/Software Deployment
Initiated
April 5, 2023
Posted
June 16, 2023
Address
5200 Illumina Way, San Diego, CA, 92122-4616

Description

NextSeq 550Dx Instrument, Catalog # 20005715, contaning Universal Copy Service software

Reason

Cybersecurity vulnerability concerning the software used for sequencing instruments.

Action

On April 5, 2023, Illumina issued a "Urgent Medical Device Recall" Notification via XXX. Illumina ask consignees to take the following actions: 1. Download & Install Software Patch: For Each Instrument Connected to the Internet: 2. Change UCS User Configuration(s): Change the UCS User configuration to standard user (with no administrator permissions) by following the instructions available in the customer communication. 3. Complete and return the Verification Form after carrying out all the steps in the instructions provided on your specific instrument(s) identified as affected in Table 1. 4. If you experience an adverse event due to this vulnerability with the use of any of the affected products, please report it to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. You can complete and submit the report online at www.fda.gov/medwatch/report.htm.

Distribution

Worldwide distribution - US Nationwide and the countries of Algeria, AU Australia, Austria, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Egypt, FR France, GB Georgia, Germany, Guatemala, Hong Kong, Hungary, IT Italy, Japan, Latvia, Lithuania, Luxembourg, Macedonia, Malta, Morocco, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, Russian Fed., Saudi Arabia, SG Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Utd. Arab Emir., Vietnam, Argentina, Aerbaijan, Belgium, Bosnia-Herz., Finland, Greece, Iceland, Ireland, Israel, Japan, Nigeria, NL Norway, Romania, Serbia, Slovakia, South Africa, T¿rkiye, Ukraine.

Quantity

1067 units